This regulation requires that anesthesia apparatus be tested at the final path to the patient following any adjustment, modification, or repair, with each gas connection verified and oxygen concentration confirmed via an oxygen analyzer before the equipment is returned to service. Defective equipment must be immediately removed from service. Service areas for oxygen equipment must be kept clean and free of flammable materials, and a scheduled maintenance program using manufacturer service manuals must be followed. This is a Life Safety Code / NFPA 99 physical plant and equipment standard applicable to facilities with anesthesia and medical gas systems.
View official sourceK924 Gas Equipment – Testing and Maintenance Requirements Anesthesia apparatus are tested at the final path to patient after any adjustment, modification or repair. Before the apparatus is returned to service, each connection is checked to verify proper gas and an oxygen analyzer is used to verify oxygen concentration. Defective equipment is immediately removed from service. Areas designated for servicing of oxygen equipment are clean and free of oil, grease, or other flammables. Manufacturer service manuals are used to maintain equipment and a scheduled maintenance program is followed. 11.4.1.3, 11.5.1.3, 11.6.2.5, 11.6.2.6 (NFPA 99) Form CMS-2786R (07/2018) Page 45 Name of Facility 2012 LIFE SAFETY CODE ID PREFIX MET NOT MET N/A REMARKS