This regulation requires facilities to test and maintain patient-care related electrical equipment (PCREE) in accordance with NFPA 99 standards, including physical integrity, resistance, leakage current, and touch current testing before initial use and after any repair or modification. Facilities must establish testing interval policies, maintain accessible service manuals and safety labels, keep records of all tests, repairs, and modifications, and ensure personnel receive ongoing training on electrical equipment use and maintenance.
View official sourceK921 Electrical Equipment – Testing and Maintenance Requirements The physical integrity, resistance, leakage current, and touch current tests for fixed and portable patient-care related electrical equipment (PCREE) is performed as required in 10.3. Testing intervals are established with policies and protocols. All PCREE used in patient care rooms is tested in accordance with 10.3.5.4 or 10.3.6 before being put into service and after any repair or modification. Any system consisting of several electrical appliances demonstrates compliance with NFPA 99 as a complete system. Service manuals, instructions, and procedures provided by the manufacturer include information as required by 10.5.3.1.1 and are considered in the development of a program for electrical equipment maintenance. Electrical equipmen t instructions and maintenance manuals are readily available, and safety labels and condensed operating instructions on the appliance are legible. A record of electrical equipment tests, repairs, and modifications is maintained for a period of time to demonstrate compliance in accordance with the facility's policy. Personnel responsible for the testing, maintenance and use of electrical appliances receive continuing training. 10.3, 10.5.2.1, 10.5.2.1.2, 10.5.2.5, 10.5.3, 10.5.6, 10.5.8