Regulation detail

K-Tag K914

Up to date
Ask Ariadne
SR
K-Tag K914 active

Electrical Systems – Maintenance and Testing

Jurisdiction: CM Agency: Centers for Medicare & Medicaid Services
PSYCH_FACILITY (60%) SNF (100%)
Plain-English summary

This regulation requires hospitals to test and maintain electrical receptacles at patient bed locations and anesthesia administration areas, including post-installation testing and interval-based retesting based on performance data. Line isolation monitors (LIMs), if present, must be tested monthly (or annually if self-testing) and after any electrical system repairs or renovations. Facilities must maintain records of all tests, repairs, and modifications, including date, location, and results. This is a life-safety/electrical systems maintenance requirement applicable to hospital settings.

View official source
Regulation text
K914 Electrical Systems – Maintenance and Testing 
Hospital-grade receptacles at patient bed locations and where deep 
sedation or general anesthesia is administered, are tested after initial 
installation, replacement or servicing. Additional testing is performed at 
intervals defined by documented performance data. Receptacles not listed 
as hospital-grade at these locations are tested at intervals not exceeding 12 
months. Line isolation monitors (LIM), if installed, are tested at intervals of 
≤ 1 month by actuating the LIM test switch per 6.3.2.6.3.6, which activates 
both visual and audible alarm. For LIM circuits with automated self-testing, 
this manual test is performed at intervals ≤ 12 months. LIM circuits are 
tested per 6.3.3.3.2 after any repair or renovation to the electric distribution 
system. Records are maintained o f required tests and associated repairs or 
modifications, containing date, room or area tested, and results. 
6.3.4 (NFPA 99)